Responsible: Effland |
| *************************************************************************************** |
| RID 1716 |
| Doc. Title | | Band 6 Cartridge - Technical Specifications | Originator: K. Saini | Date: 2007-09-18 |
| Description: | | (Review Item Question) |
| |
| Although the changes suggested in a related CRE have not yet been included into the Band |
| 6 cartridge specifications, the Band 6 cartridge group should highlight any significant |
| problems/issues with respect to these proposed changes for the benefit of the CDR review |
| panel. |
| |
| FEND-40.02.06.00-154-A-CRE at |
| http://edm.alma.cl/forums/alma/dispatch.cgi/f.frontendipt/docProfile/100574/ |
Suggested Solution: |
| *************************************************************************************** |
| RID 1717 |
| Doc. Title | | Band 6 Cartrdige Design Report | Originator: K. Saini | Date: 2007-09-18 |
| Description: | | (Review Item Discrepancy) |
| |
| Section 9 - Serial numbers 3.3.2.2 and 3.7.4 |
| |
| Please confirm if these non compliances are generic or article specific. If generic, then |
| please indicate how these will be resolved (modification of design, CRE etc) |
Suggested Solution: |
| *************************************************************************************** |
| RID 1718 |
| Doc. Title | | Band 6 Cartrdige Design Report | Originator: K. Saini | Date: 2007-09-18 |
| Description: | | (Review Item Comment) |
| |
| Section 7.7 - Table 10 |
| |
| The actual physical heat loads seem to be okay, but the "calculated" entries in Table 10 |
| seem to be for the dynamic heat load, while the "allowed" numbers (compliant but slightly |
| different from numbers in the ICD) appear to be for the total allocation (static plus |
| dynamic). |
Suggested Solution: |
------------------------------------------------------------------------------------------------------------- |
Printed 2007-09-18 16:52:46 | Page 5 of 8 |