next up previous contents index external
Next: Self-calibration criteria Up: Calibration Details Previous: VLA flux density calibration

VLA polarization calibration

         

Instrumental

To calibrate the instrumental polarizations, observe one unresolved calibrator, whether polarized or not, at least three times, more often if possible. Distribute these observations to cover a range in parallactic angle tex2html_wrap_inline3063 of tex2html_wrap_inline3065 : this will separate any polarization of the calibrator from the required instrumental terms (see Lecture 7). Programs involving long ( tex2html_wrap_inline3119 4 hr) syntheses of single sources will normally be able to derive the instrumental polarization calibration from the observations of the external gain calibrator. When determining the integration time for the instrumental polarization calibration, bear in mind that the cross-polarization (leakage) terms whose relative amplitudes and phases are to be determined will likely produce false polarized intensities that are only a few percent of the flux density of the calibrator. The instrumental polarization calibration must be done at each frequency for which polarimetry is required. The most efficient way to do this is to cycle through the frequencies used for the target observations each time you point the array at the calibrator. If you omit an instrumental polarization calibration, you will be unable to determine small degrees of polarization reliably, or to deconvolve polarized extended structures properly. (Antenna-to-antenna polarization differences distort the polarization images in ways that do not satisfy the convolution theorem.)

Position Angle

To calibrate the polarization position angle scale, observe 3C286 or 3C138 at least once during your observing run at each relevant frequency. You will determine the apparent position angles of the linear polarization of these sources after you have finished observing and after calibrating the total intensity data. The difference between the apparent and the nominal values of these position angles is corrected by adjusting the phase difference between the left and right circular polarizations, using a procedure that is described in detail in the AIPS Cookbook. It is advisable to alert the array operator to the presence of this calibration in your program, so that the observations of 3C286 or 3C138 can be extended or rescheduled if necessary to avoid losing them to equipment failure. Note that this calibration is essential if you wish to use your polarization position angle data, e.g. to determine magnetic field directions or to measure the Faraday rotation of a source.

Ionospheric Faraday Rotation

At wavelengths of 18 cm and longer, the position angle calibration may appear to be time variable because of fluctuations in the ionospheric Faraday rotation (Lecture 7). If you will use the polarization position angle information at long wavelengths, it is worth monitoring one polarized calibrator in the same part of the sky as your target(s) throughout your observing, to see if its apparent position angle changes significantly. If this calibration shows that the ionospheric changes are less than about tex2html_wrap_inline2659 , it may be satisfactory to interpolate the observed position angle changes as a function of time when you set the position angle scale by adjusting the relative phase of the left and right circularly polarized channels. If you see larger changes, especially in a calibrator some tens of degrees from your target, it may be impossible to correct for them accurately without using an ionospheric model. Unfortunately, a source of critical-frequency data that was used to estimate ionospheric models for many years at the VLA (via the FARAD programs) is no longer available. Until it is replaced (possibly by data from GPS systems), there is no way to repair VLA polarimetric data sets in which the ionospheric Faraday rotations are large. Observing a polarized calibrator can therefore be a ``warning light'' for such ionospheric Faraday rotation problems, but it does not guarantee a way to correct them. Fortunately, ionospheric effects are normally negligible above 4 GHz.


next up previous contents index external
Next: Self-calibration criteria Up: Calibration Details Previous: VLA flux density calibration

abridle@nrao.edu
Thu Jul 11 16:26:53 EDT 1996