Next: Rationale for Band 1
Up: An Infrared Water Vapour
Previous: Future plans
The theoretical calculations presented above suggest the infrared water
vapour monitor (IWVM) should be able to
achieve a 1 sigma sensitivity of 1
m pwv in 1 second. This sensitivity
is appropriate for an average pwv of 1 mm; the sensitivity will be 30
percent better when the pwv is 0.5 mm. Now, according to
Lay et al. (MMA memo 209), 1
m of pwv corresponds to 6
m in
the optical path. The currently recommended design goal for ALMA is
an accuracy of 10(1+w)
m in optical path per baseline, where
w is the atmospheric precipitable water vapour (in millimeters).
Since the path
difference is determined from measurements at two telescopes, this means
the accuracy needed at each telescope is about 14
m of path or 2
m of pwv when the total pwv is 1 mm.
Thus, if the IWVM can achieve the predicted theoretical
sensitivity, it should be able to provide the required accuracy.
The preliminary analysis of the results from the first run with
the IWVM gives a 1 sigma accuracy of 10
m pwv (60
m of path)
in one second of integration.
In comparison, the uncooled 183 GHz radiometers on the JCMT and the CSO
achieved a 1 sigma accuracy of 0.1-0.2 K in 10 seconds
(Wiedner & Hills, Imaging 99 proceedings). From
the calibration on a day when the pwv was 2.2 mm, this rms
corresponds to a 1 sigma accuracy of 20-40
m in path.
In comparison, if we were to average
the IWVM data over 10 seconds instead of 1 second, the sensitivity
from the first run would be 3
m pwv or 20
m in path. Thus,
the current IWVM and the current 183 GHz radiometers likely have similar
sensitivities.
The expected rms for the existing 183 GHz radiometers from just thermal
noise is about 0.06 K; there were substantial fluctuations in the
instrumental gain which brought the noise up (Wiedner & Hills 1999).
Also, the current plan for ALMA is to cool the 183 GHz radiometers,
which will bring the system temperature down by a large factor.
As an example, if the system temperature of the uncooled 183 GHz system is
2500 K and that of the cooled system is 400 K, then the thermal noise
would drop from 0.06 K to 0.01 K in a 10 second integration or
0.033 K in a one second integration. This means the one sigma accuracy of a
cooled 183 GHz radiometer in one second of integration
would be 7
m in path or about 1
m in pwv.
However, this sensitivity is appropriate for fairly wet conditions
(2.2 mm pwv; 3.9 K/turn or 4.6 K/mm path). From Figure 2 in Wiedner & Hills
(1999), when the atmosphere is very dry (0.4 mm pwv),
the sensitivity at the line center rises to 40 K/mm path. If the 183 GHz
radiometers for ALMA can measure close to the line center, the sensitivity
under these conditions would be 0.8
m in path or 0.13
m
in pwv. However, at these levels, the sensitivity would likely be
dominated by other factors such as gain stability, uncertainty in the
altitude of the water vapour, etc. (see Lay et al., MMA memo 209).
In comparison, the expected sensitivity of the IWVM is 4
m in path
(0.66
m in pwv) with an atmosphere of 0.5 mm pwv and 8.5
m in path
with an atmosphere of 2 mm pwv. (Note that these sensitivities are
estimated for Mauna Kea; the sensitivity should be even better at an
altitude of 5000 m.) Thus, if the IWVM can achieve the
theoretically predicted sensitivity, it could be competitive with the
183 GHz radiometers and should be able to meet the design goals for ALMA.
Figure F1:
IWVM Sensitivity versus Water Vapor
|
Figure F2:
Picture of system on the JCMT
|
Figure F3:
Theoretical Curve of Growth
|
Figure F4:
Stability of blackbody reference measurements
|
Figure F5:
Three sky-dips during a 50 scan sequence
|
Figure F6:
Similar to Fig. F5, but on a dry and stable night
|
Figure F7:
Early results of algorithm
|
Figure F8:
Comparison of 183 GHz (lower trace) and infrared system
|
Next: Rationale for Band 1
Up: An Infrared Water Vapour
Previous: Future plans
Al Wootten
2000-04-04